Alignment of a tiled-grating compressor in a high-power chirped-pulse amplification laser system |
| |
Authors: | Hornung Marco Bödefeld Ragnar Siebold Mathias Schnepp Matthias Hein Joachim Sauerbrey Roland Kaluza Malte C |
| |
Affiliation: | Institut für Optik und Quantenelektronik, Friedrich-Schiller-Universit?t Jena, Germany. hornung@ioq.uni-jena.de |
| |
Abstract: | We present a novel technique to align a tiled grating in all five relevant degrees of freedom utilized in the compressor of the high-power chirped-pulse amplification laser system POLARIS at the Institute for Optics and Quantum Electronics, Jena, Germany. With this technique, alignment errors of the two gratings with respect to each other can be detected with an accuracy of 1 microrad for the rotational and 40 nm for the translational degrees of freedom. This is well sufficient to recompress 1030 nm pulses, which were stretched to 2.2 ns before amplification, to their bandwith limit of 150 fs. |
| |