首页 | 本学科首页   官方微博 | 高级检索  
     


Wavelength-tuning interferometry of intraocular distances
Authors:Lexer F  Hitzenberger C K  Fercher A F  Kulhavy M
Abstract:We describe basic principles of wavelength-tuning interferometry and demonstrate its application in ophthalmology. The advantage of this technique compared with conventional low-coherence interferometry ranging is the simultaneous measurement of the object structure without the need for a moving reference mirror. Shifting the wavelength of an external-cavity tunable laser diode causes intensity oscillations in the interference pattern of light beams remitted from the intraocular structure. A Fourier transform of the corresponding wave-number-dependent photodetector signal yields the distribution of the scattering potential along the light beam illuminating the eye. We use an external interferometer to linearize the wave-number axis. We obtain high resolution in a model eye by slow tuning over a wide wavelength range. With lower resolution we demonstrate the simultaneous measurement of anterior segment length, vitreous chamber depth, and axial eye length in human eyes in vivo with data-acquisition times in the millisecond range.
Keywords:
本文献已被 PubMed 等数据库收录!