首页 | 本学科首页   官方微博 | 高级检索  
     


Development and experimental evaluation of aerodynamic lens as an aerosol inlet of single particle mass spectrometry
Authors:Kwang-Sung Lee  Sung-Woo Cho  Donggeun Lee
Affiliation:School of Mechanical Engineering, Pusan National University, Busan 609-735, Republic of Korea
Abstract:This study described a development and an experimental evaluation of an efficient aerodynamic lens inlet of the single particle mass spectrometry. Several key designing parameters and systematic factors were investigated for the whole lens system through a full numerical simulation. From many tests for various designs of the system, we showed that Mach number was not an independent parameter but interrelated well with flow Reynolds numbers and pressures upstream of the orifices. By manipulating the parameters, we showed for the first time a possibility that there exist a universal correlation between optimal Stokes number and a new factor incorporating the other dimensionless variables and a design parameter. The universality was confirmed by the full simulation results. We demonstrated that the new design of the system was capable of focusing ultrafine aerosols in the size range of 30–700 nm. At two different operating conditions, the formations of sub-millimeter beams of 30–300 nm NaCl aerosols are verified by light scattering imaging as well as microprobe observation of deposited aerosol beams. Finally, the measured sizes of aerosol beams agree reasonably well with those from the simulations as a function of particle size.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏