The validity of using a video-based motion analysis system for measuring maximal area of fingertip motion and angular variation
Authors:Su F-C  Kuo L-C  Chiu H-Y  Hsu H-Y
Affiliation:Institute of Biomedical Engineering, National Cheng Kung University, Tainan, Taiwan.
Abstract:The aim of the study was to verify the application of a three-dimensional video motion analysis system to evaluate maximal fingertip motion area and angular variation of the hand by comparison and correlation with videofluoroscopic analysis. Eight normal subjects were recruited in this study. The maximal motion area of the fingertip and the angles of the metacarpal phalangeal (MP), proximal interphalangeal (PIP) and distal interphalangeal (DIP) joints in performing five sequential postures for functional evaluation of the hand were measured using a video motion analysis system and a fluoroscopy system respectively. The results indicated that the intraclass correlation coefficient (ICC) of the calculated maximal fingertip motion area between the two methods was 0.9597. The ICC for total active motion (TAM) measurements of three finger joints was 0.940 between the surface and bony landmarks by fluoroscopy, 0.952 between the surface landmarks from fluoroscopy and motion analysis, and 0.927 between the bony landmark from fluoroscopy and surface markers from motion analysis. The ICC for angular measurements between three different paired assessments was 0.9650, 0.8896 and 0.8799 for the MP, PIP and DIP joints respectively. The results indicate that motion analysis is a practical method for assessing impairment of the hand.