首页 | 本学科首页   官方微博 | 高级检索  
     


Construction and tests of a heart scanner based on superconducting sensors cooled by small stirling cryocoolers
Authors:A P Rijpma  C J H A Blom  A P Balena  E de Vries  H J Holland  H J M ter Brake  H Rogalla
Affiliation:a Low Temperature Division, Department of Applied Physics, University of Twente, P.O. Box 217, 7500 AE Enschede, Netherlands;b Philips Medical Systems Nederland B.V., Afd. MR Scan Software, P.O. Box 10000, 5680 DA Best, Netherlands;c Hollandse Signaalapparaten B.V., P.O. Box 42, 7550 GD Hengelo, Netherlands
Abstract:At the University of Twente, a heart scanner has been designed and constructed that uses superconducting devices (superconducting quantum interference devices (SQUIDs)) to measure the magnetic field of the heart. A key feature is the elimination of liquid cryogens by incorporating cryocoolers. In the design, two coolers are operated in counter-phase to reduce the mechanical interference. In addition to the application of ferromagnetic shields around the compressors, the magnetic cooler interference is reduced by placing the SQUID magnetometers coplanar with respect to the coolers. In this way, the cooler noise was reduced to a level below the intrinsic sensor noise: 0.16 pT/√Hz. A temperature of 60 K was realised with a cool-down time of about 2 h. The corresponding heat load to the coolers is roughly 0.9 W. Magnetocardiograms were recorded inside a magnetically shielded room.
Keywords:Cryocooler  SQUID  Magnetic interference  Stirling cooler  Superconductivity
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号