A procedure to determine the correct thickness of an object with confocal microscopy in case of refractive index mismatch
Authors:L C KUYPERS  W F DECRAEMER  J J J DIRCKX  & J-P TIMMERMANS
Affiliation:Laboratory of Biomedical Physics, Department of Physics, and;Laboratory of Cell Biology and Histology, Department of Biomedical Sciences, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerpen, Belgium
Abstract:A difference in refractive index (n) between immersion medium and specimen results in increasing loss of intensity and resolution with increasing focal depth and in an incorrect axial scaling in images of a confocal microscope. Axial thickness measurements of an object on such images are therefore not exact. The present paper describes a simple procedure to determine the correct axial thickness of an object with confocal fluorescence microscopy. We study this procedure for a specimen that has a higher refractive index than the immersion medium and with a thickness up to 100 µm. The measuring method was experimentally tested by comparing the thickness of polymer layers measured on axial images of a confocal microscope in case of a water–polymer mismatch to reference values obtained from an independent technique, i.e. scanning electron microscopy. The case when the specimen has a lower refractive index than the immersion medium is also shown by way of illustration. Measured thickness data of a water layer and an oil layer with the same actual thickness were obtained using an oil-immersion objective lens with confocal microscopy. Good agreement between theory and experiment was found in both cases, consolidating our method.
Keywords:Axial scaling  confocal  focal shift  intensity drop                n-mismatch  point spread function  polymer film  quantitative microscopy  thickness
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号