首页 | 本学科首页   官方微博 | 高级检索  
     


Usefulness of non-lead aprons in radiation protection for physicians performing interventional procedures
Authors:Zuguchi Masayuki  Chida Koichi  Taura Masaaki  Inaba Yohei  Ebata Ayako  Yamada Shogo
Affiliation:Course of Radiological Technology, Health Sciences, Tohoku University Graduate School of Medicine, 2-1 Seiryo, Aoba, Sendai 980-8575, Japan. zuguchi@wm.pdx.ne.jp
Abstract:
At present, interventional radiology (IVR) tends to involve long procedures (long radiation duration), and physicians are near to the source of scattered radiation. Hence, shielding is critical in protecting physicians from radiation. Protective aprons and additional lead-shielding devices, such as tableside lead drapes, are important means of protecting the physician from scattered radiation. The purpose of this study was to evaluate whether non-lead aprons are effective in protecting physicians from radiation during IVR procedures. In this study, the radiation protection effects of commercially available protective lead and non-lead aprons, when exposed to diagnostic X rays, are compared. The performance of these non-lead and lead aprons was similar for scattered X rays at tube voltages of 60-120 kV. Properly designed non-lead aprons are thus more suitable for physicians because they weigh approximately 20% less than the lead aprons, and are non-toxic.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号