首页 | 本学科首页   官方微博 | 高级检索  
     


Charge collection efficiency characterization of a HgI2 Frisch collar spectrometer with collimated high energy gamma rays
Authors:A. Kargar  E. AriesantiS. James  D.S. McGregor
Affiliation:a Radiation Monitoring Devices, Inc., 44 Hunt Street, Watertown, MA 02472, USA
b S.M.A.R.T. Laboratory, Mechanical and Nuclear Engineering Department, Kansas State University, Manhattan, KS 66506, USA
Abstract:In this study, a 2.1×2.1×4.1 mm3 HgI2 Frisch collar device was characterized through probing the device with a highly collimated 662 keV gamma rays (137Cs check source) along the length and width of the device. In a systematic series of experiments, the detector was probed along its central line under different operating voltages of 1600, 1300, 1000, 800, 600 and 500 V. The experimental results were confirmed through a simulation of the charge collection for a device with the same size and operating conditions. It is shown that the HgI2 Frisch collar device has a uniform response to gamma rays over two-thirds of the detector volume. The HgI2 crystals and the Frisch collar detectors were grown and fabricated within the S.M.A.R.T Laboratory at Kansas State University.
Keywords:HgI2   Frisch collar detector   Gamma-ray
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏