烟气超声流量计时间测量准确度校准方法研究
引用本文:樊洁云,张亮,方立德,王池,刘洋.烟气超声流量计时间测量准确度校准方法研究[J].中国测试,2020(2):28-33.
作者姓名:樊洁云  张亮  方立德  王池  刘洋
作者单位:;1.河北大学质量技术监督学院;2.中国计量科学研究院;3.郑州计量先进技术研究院;4.河南省环境监测技术重点实验室
基金项目:国家重点研发计划(2016YFF0204403)
摘    要:大口径气体超声波流量计广泛应用于烟道排放监测和工业控制等领域。超声波流量计的传播时间测量准确度校准是非实流校准的重要环节,通过建立超声波流量计时间测量准确度校准装置,实现对大口径气体超声流量计传播时间的非实流校准。该文首先提出3种不同的超声流量计传播时间校准方法,分析不同方法的影响因素,其次通过改变探头间距离,对不同探头间距时超声波流量计传播时间的测量误差进行校准。试验结果表明:超声波流量计传播时间的测量误差随距离变化,并确定利用标准声速对超声波流量计传播时间进行修正的方法更为准确,测量结果不确定度为0.2%。

关 键 词:超声流量计  传播时间  校准方法  不确定度

Study on calibration method of time measurement accuracy of flue gas ultrasonic flowmeter
Authors:FAN Jieyun  ZHANG Liang  FANG Lide  WANG Chi  LIU Yang
Affiliation:(College of Quality and Technical Supervision,Hebei University,Baoding 071002,China;National Institute of Metrology,Beijing 100029,China;Zhengzhou Institute of Metrology,Zhengzhou 450001,China;Henan Key Laboratory of Environmental Monitoring Technology,Zhengzhou 450004,China;Henan Ecological Environmental Monitoring Center,Zhengzhou 450004,China)
Abstract:Large diameter gas ultrasonic flowmeter is widely used in stack emission monitoring and industrial control.The accuracy of the propagation time measurement of the ultrasonic flowmeter is an important part of dry calibration.Through the establishment of the ultrasonic flowmeter time measurement accuracy calibration facility,it can realize the dry calibration of the propagation time of large diameter gas ultrasonic flowmeter.This paper firstly proposes three different methods of ultrasonic flowmeter propagation time calibration,and analyzes the influencing factors of different methods,and then calibrates the measurement error of ultrasonic flowmeter propagation time when different probe spacing is changed.The test results show that the measurement error of ultrasonic flowmeter propagation time varies with distance,and it is more accurate to determine the propagation time of the ultrasonic flowmeter using the standard sound velocity.The uncertainty of the measurement result is 0.2%.
Keywords:ultrasonic flowmeter  transit time  calibration method  uncertainty