Laminar and blazed type holographic gratings for a versatile soft x-ray spectrograph attached to an electron microscope and their evaluation in the 50-200 eV range
Authors:Imazono Takashi  Koike Masato  Kawachi Tetsuya  Hasegawa Noboru  Koeda Masaru  Nagano Tetsuya  Sasai Hiroyuki  Oue Yuki  Yonezawa Zeno  Kuramoto Satoshi  Terauchi Masami  Takahashi Hideyuki  Handa Nobuo  Murano Takanori  Sano Kazuo
Affiliation:Quantum Beam Science Directorate, Japan Atomic Energy Agency, Kizugawa, Kyoto, Japan. imazono.takashi@jaea.go.jp
Abstract:Laminar and blazed type holographic varied-line-spacing spherical gratings for use in a versatile soft x-ray flat-field spectrograph attached to an electron microscope are designed, fabricated, and evaluated. The absolute diffraction efficiencies of laminar (or blazed) master and replica gratings at 86.00° incidence evaluated by synchrotron radiation show over 5% (or 8%) in the 50-200 eV range with the maxima of 22% (or 26%-27%). Also the resolving power evaluated by a laser produced plasma source is in excess of 700 at the energy near the K emission spectrum of lithium (~55 eV) for all gratings. Moreover, the K emission spectrum of metallic Li with high spectral resolution is successfully observed with the spectrograph attached to a transmission electron microscope.