首页 | 本学科首页   官方微博 | 高级检索  
     


Spatial-resolution enhancement in computed tomography
Authors:Pan Xiaochuan  Yu Lifeng  Kao Chien-Min
Affiliation:Department of Radiology, The University of Chicago, Chicago, IL 60637, USA. xpan@uchicago.edu
Abstract:We propose an approach that combines an asymmetric fan-beam configuration and a new reconstruction algorithm to enhancing spatial resolution in computed tomography (CT). The asymmetric configuration can be achieved by changing the center of rotation (COR) from the conventional symmetric configuration. It does not, however, require new detectors and X-ray source nor alter the relative geometry between the detector and the X-ray source. By effectively reducing the distance of the COR to the X-ray source, the asymmetric configuration can increase the effective sampling density in projection data without reducing the size of the field of view (FOV). The proposed algorithm, on the other hand, can reconstruct images directly from data acquired with this asymmetric configuration. We performed numerical studies to demonstrate and validate the proposed acquisition/reconstruction approach. Results in these studies confirm that the proposed approach can lead to enhanced spatial resolution in reconstructed images. The proposed acquisition/reconstruction approach may find applications in micro-CT and industrial CT in which the CORs may be changed.
Keywords:
本文献已被 PubMed 等数据库收录!