首页 | 本学科首页   官方微博 | 高级检索  
     

TOF-PET探测器定时性能初步研究
引用本文:刘军辉,李道武,章志明,廖燕飞,黄先超,王培林,陈研,贠明凯,丰宝桐,帅磊,王宝义,张天保,魏龙.TOF-PET探测器定时性能初步研究[J].原子能科学技术,2011,45(12):1522-1526.
作者姓名:刘军辉  李道武  章志明  廖燕飞  黄先超  王培林  陈研  贠明凯  丰宝桐  帅磊  王宝义  张天保  魏龙
作者单位:1.中国科学院 ;高能物理研究所 ;核分析技术重点实验室,北京100049;2.北京市射线成像技术与装备工程技术研究中心,北京100049;3.中国科学院 ;研究生院,北京100049
基金项目:国家自然科学基金资助项目
摘    要:本工作采用LYSO晶体阵列和平板型位置灵敏光电倍增管H8500对TOF-PET探测器进行了初步设计,能分辨3.2mm×3.2mm的晶体阵列。与BaF2探测器组成时间谱仪,测量正电子湮没射线的瞬发时间谱,得到379ps的符合时间分辨率。根据已测量的BaF2探测器时间分辨率(159ps),可计算得到所设计的TOF-PET探测器时间分辨率为344ps。预计与此相同的两个TOF-PET探测器对正电子湮没射线的符合时间分辨率可达486ps左右,为新型PET的研究提供了有利条件。

关 键 词:正电子发射断层扫描仪    飞行时间    硅酸镥    H8500

Pilot Study of Timing Performance for Time-of-Flight Positron Emission Tomography Detector
LIU Jun-hui,LI Dao-wu,ZHANG Zhi-ming,LIAO Yan-fei,HUANG Xian-chao,WANG Pei-lin,CHEN Yan,YUN Ming-kai,FENG Bao-tong,SHUAI Lei,WANG Bao-yi,ZHANG Tian-bao,WEI Long.Pilot Study of Timing Performance for Time-of-Flight Positron Emission Tomography Detector[J].Atomic Energy Science and Technology,2011,45(12):1522-1526.
Authors:LIU Jun-hui  LI Dao-wu  ZHANG Zhi-ming  LIAO Yan-fei  HUANG Xian-chao  WANG Pei-lin  CHEN Yan  YUN Ming-kai  FENG Bao-tong  SHUAI Lei  WANG Bao-yi  ZHANG Tian-bao  WEI Long
Affiliation:1.Key Laboratory of Nuclear Analyses Technology, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China;2.Beijing Engineering Research Center of Radiographic Techniques and Equipments, Beijing 100049, China;3.Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:In this work, LYSO scintillators and position sensitive photomultiplier tube (PS-PMT) were used to design the TOF-PET detector, and LYSO crystals of 3.2 mm×3.2 mm pitch could be distinguished. The coincidence time resolution of 379 ps was measured for the γ rays produced by positron annihilation with a reference detector of BaF2, whose time resolution for 511 keV γ ray was measured as 159 ps. By subtracting the contribution of the BaF2 detector, 344 ps time resolution could be calculated for the designed TOF-PET detector, and 486 ps coincidence time resolution for γ rays of positron annihilation can be expected for two such detectors. It means that the new detector for TOF-PET is feasible.
Keywords:positron emission tomography  time-of-flight  LYSO  H8500  
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《原子能科学技术》浏览原始摘要信息
点击此处可从《原子能科学技术》下载全文