Quantitative comparison between full-spectrum and filter-based imaging in hyperspectral fluorescence microscopy
Authors:Gao L  Hagen N  Tkaczyk T S
Affiliation:Department of Bioengineering, Rice University, 6100 Main Street, Houston, TX 77005, USA.
Abstract:We implement a filterless illumination scheme on a hyperspectral fluorescence microscope to achieve full-range spectral imaging. The microscope employs polarisation filtering, spatial filtering and spectral unmixing filtering to replace the role of traditional filters. Quantitative comparisons between full-spectrum and filter-based microscopy are provided in the context of signal dynamic range and accuracy of measured fluorophores' emission spectra. To show potential applications, a five-colour cell immunofluorescence imaging experiment is theoretically simulated. Simulation results indicate that the use of proposed full-spectrum imaging technique may result in three times improvement in signal dynamic range compared to that can be achieved in the filter-based imaging.
Keywords:Hyperspectral fluorescence microscopy  filterless illumination  spectroscopy