首页 | 本学科首页   官方微博 | 高级检索  
     


Scanning tomographic acoustic microscopy using shear waves
Authors:Ko D  Meyyappan A
Affiliation:Dept. of Electron. Eng., Mok-Won Univ., Taejon.
Abstract:
We propose to use shear waves instead of longitudinal waves in a novel scanning tomographic acoustic microscope (STAM) in which the specimens are solid. When a specimen with a shear modulus is immersed in the microscope's water bath, mode conversion takes place at the water-solid interface. The shear wave energy is detectable and can be used for image reconstruction. Although wave transmission in most solid specimens is limited to about 20 degrees for longitudinal waves, it is about twice that for shear waves. Also, velocities of shear waves are lower than those of longitudinal waves and hence the wavelengths at the same frequency are smaller. For these and other reasons we can expect that for many specimens the resolution of a shear-wave STAM to be substantially better than that of a longitudinal-wave STAM. We use computer simulation in order to compare the operation of a shear-wave STAM with that of the conventional longitudinal-wave STAM. We have simulated tomographic reconstruction for each. The corresponding critical angles of incidence are computed and tomographic reconstructions of a particular solid specimen is obtained by using the back-and-forth propagation algorithm (BFP). Our simulation results show that shear-wave STAM has better resolution than longitudinal-wave STAM.
Keywords:
本文献已被 PubMed 等数据库收录!