首页 | 本学科首页   官方微博 | 高级检索  
     

极端气温永磁体磁通测试系统
引用本文:王顺杰,高伟波.极端气温永磁体磁通测试系统[J].磁性材料及器件,2017,48(4).
作者姓名:王顺杰  高伟波
作者单位:宁波市计量测试研究院 国家磁性材料产业计量测试中心(筹),浙江宁波,315048
摘    要:设计搭建了一台永磁体极端气温磁通测试系统。系统主要由温度控制、磁性能测试与连接组件三部分组成。通过合理的结构设计实现了可变温度环境下磁通的准确测试。利用本测试系统对两种典型永磁样品(烧结钕铁硼、永磁铁氧体)进行测试。实验结果表明:在-60~100℃温度区间内,样品的开路磁通随着温度的升高线性衰减。牌号N50M烧结钕铁硼的磁通平均温度系数为-0.09%/℃,牌号Y25永磁铁氧体的磁通平均温度系数为-0.15%/℃。

关 键 词:永磁体  极端气温  磁通测试

Design of flux measurement system for permanent magnet at extremely low temperature
WANG Shun-jie,GAO Wei-bo.Design of flux measurement system for permanent magnet at extremely low temperature[J].Journal of Magnetic Materials and Devices,2017,48(4).
Authors:WANG Shun-jie  GAO Wei-bo
Abstract:A system was designed and fabricated for measuring magnetic flux at extremely low temperature. The equipment consists of three parts: temperature control, magnetic measurement and connector. Precise flux measuring at variable temperature can be realized by reasonable structure design. Nd-Fe-B sintered magnet and permanent ferrite are two typical permanent magnet materials, of which grade N50M and Y25 samples were measured with the system at different temperatures. The test results show that, between the temperature ranging from-60℃ to 100℃, the magnetic flux linearly decreases as the temperature increases. The flux temperature coefficient of Nd-Fe-B sintered magnet and permanent magnetic ferrite was-0.09%/℃ and-0.15%/℃, respectively.
Keywords:permanent magnet  extreme temperature  magnetic flux measurement
本文献已被 CNKI 万方数据 等数据库收录!