首页 | 本学科首页   官方微博 | 高级检索  
     


Stability of proximal femoral grafts in canine hip arthroplasty
Authors:JP Heiner  SS Kohles  PA Manley  R Vanderby  MD Markel
Affiliation:Division of Orthopedic Surgery, University of Wisconsin, Madison 53792-3228, USA.
Abstract:In a canine model, the fixation stability of a prosthesis and proximal bone graft composite were measured relative to the distal femur. One group had the prosthesis graft composite cemented into the distal femur. The second group had the prosthesis graft composite press fit into the distal femur for biologic ingrowth. Displacements of the proximal femoral grafts relative to the host bone in each group were measured after ex vivo (acute with graft) implantation and 4 months after implantation. A third group with no osteotomy (acute intact) simulated perfect graft to host bone union. Relative displacements representing 6 degrees freedom (translation and rotation) were calculated from the displacement values measured by 9 eddy current transducers. Measurements of displacement were used to test the hypothesis that distal press fit fixation equals distal cement fixation at 4 months after implantation. In all cases the measured translations and rotations of the graft to implant construct were small and of a magnitude that should encourage bone ingrowth (< 0.05 mm and < 0.1 degree, respectively). The stability of the press fit group at 4 months was not significantly different from the cemented group in axial and transverse displacement during axial and transverse loading, respectively. There was no difference in stabilities at 4 months between distal press fit and cemented fixation in hip replacements requiring a proximal femoral graft.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号