Changes of intraaortic balloon volume during pumping in a mock circulation system |
| |
Authors: | T Sakamoto H Akamatsu MT Swartz Y Shoji S Kazama K Fujiyoshi Y Hashiguchi A Suzuki M Miyashita H Arai A Suzuki |
| |
Affiliation: | Department of Cardiothoracic Surgery, School of Medicine, Tokyo Medical and Dental University, Japan. |
| |
Abstract: | The effects of hemodynamic changes on intraaortic balloon (IAB) volumes were studied experimentally using a helium tachometer in a mock circulatory system. The IAB volume decreased with increases in the heart rate and mean aortic pressure, but the degree of reduction in the IAB volume was different among 4 commercially available IAB drivers. Improvements in IAB consoles are needed to compliment the progress made in percutaneous techniques for IAB insertion resulting in smaller balloons. |
| |