首页 | 本学科首页   官方微博 | 高级检索  
     


The influence of misalignment on the uncertainty of vertical torque calibration machine
Authors:K. M. Khaled  G. Aggag  A. E. Abuelezz  M. G. Elsherbiny
Affiliation:(1) Dept. Movement Sciences, Nutrition and Toxicology Research Institute Maastricht, Maastricht University, P.O. Box 616, 6200 MD Maastricht, The Netherlands;(2) Dept. of Biomedical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands;(3) Dept. of Design and Engineering Facility, Eindhoven University of Technology, Eindhoven, The Netherlands
Abstract:Misalignment of torque calibration machines can cause parasitic bending loads inside the torque transducers in use. These parasitic loads may affect the machine uncertainty, specially for secondary torque calibration machines. Therefore, this study discusses the contribution of misalignment to the measurement uncertainty for a newly designed 3 kNm secondary vertical torque calibration machine. Radial axes shift up to 4 mm was applied which introduce small increased uncertainty.
Keywords:
本文献已被 SpringerLink 等数据库收录!