Bilateral Comparison of Aluminum Fixed-Point Cells Using Standard Platinum Resistance Thermometer
Authors:J. Bojkovski  A. Peruzzi  R. Bosma  V. Batagelj
Affiliation:1. Metrology Institute of the Republic of Slovenia/University of Ljubljana-Faculty of Electrical Engineering/Laboratory of Metrology and Quality (MIRS/UL-FE/LMK), Ljubljana, Slovenia
2. VSL, Delft, The Netherlands
Abstract:The objective of Project EURAMET 1114 (Bilateral comparison of a freezing point of aluminum) in the field of thermometry is to compare realization of a freezing point of aluminum (660.323???C) between the Dutch national laboratory VSL and the Slovenian national laboratory MIRS/UL-FE/LMK using a long-stem 25 ?? standard platinum resistance thermometer (SPRT). Both laboratories had participated in a number of inter-comparisons on the level of EURAMET and also on BIPM CCT level (VSL). MIRS/UL-FE/LMK laboratory recently acquired a new fixed-point cell which had to be validated in the process of intercomparison. A quartz-sheathed SPRT was selected and calibrated at MIRS/UL-FE/LMK at the aluminum freezing point and at the water triple point. A second set of measurements was made on the same SPRT and at the same fixed points at VSL (NL). After its return, the SPRT was again recalibrated at MIRS/UL-FE/LMK. In the comparison the W value of the SPRT was used. The results of the internal and external intercomparisons confirmed that the new aluminum cell of the MIRS/UL/FE-LMK realizes a temperature that agrees with the VSL aluminum fixed point within the uncertainty limits of both laboratories. Furthermore, the results of this bilateral-comparison were compared with results that both laboratories achieved in the EURAMET K4 (Project 820) and were found to be in agreement.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号