Doppler angle and flow velocity estimations using the classic and transverse Doppler effects |
| |
Authors: | Shi B R Chiang H K Kuo C D Lin W L Lee S K |
| |
Affiliation: | Inst. of Biomed. Eng., Nat. Yang-Min Univ., Taipei. |
| |
Abstract: | ![]() Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, the authors developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. The authors have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45 degrees and 80 degrees Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5 degrees . This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|