Aquakem 200离散分析仪测定水中氨氮的研究
引用本文:曾次元,任嫣婵,赵鉴,洪华成.Aquakem 200离散分析仪测定水中氨氮的研究[J].城市公用事业,2010,24(4):31-33.
作者姓名:曾次元  任嫣婵  赵鉴  洪华成
作者单位:上海市供水调度监测中心
摘    要:对采用Aquakem 200离散分析仪测定水中的氨氮(水杨酸法),该文研究了方法的检出限、精密度、准确度及水样加标回收率。结果显示:方法检出限为0.003mg/L,0.200mg/L和1.00mg/L标准溶液的相对标准偏差分别为0.59%和0.68%,考核样测量值1.51mg/L,在标准值1.50±0.07mg/L范围内,水样加标回收率在94%~102%之间。仪器操作简单,分析速度快,尤其适合大批量水样中氨氮浓度的测定。

关 键 词:Aquakem  200离散分析仪  氨氮  检测

Determination of Ammonia Nitrogen in Water by Aquakem 200 Discrete Analyzer
Abstract:By using Aquakem 200 discrete analyzer to determine ammonia nitrogen in water, the detection limit, precision, accuracy, recovery of standard addition are investigated. The results show that the detection limit is 0.003 mg/L. The relative standard deviation of 0.200 mg/L and 1.00 mg/L ammonia nitrogen standard solution is 0.59 % and 0.68 % respectively. The determination result of QC sample is 1.51 mg/L, within the range of assigned value 1.50 ± 0.07 mg/L. The recovery of practical samples is between 94 % and 102 %. Due to easy operation and quick analysis, it is suitable for the determination of ammonia nitrogen in water, especially for large quantity of samples.