首页 | 本学科首页   官方微博 | 高级检索  
     


A simple engineering calibration for platinum resistance thermometers in the temperature range 4.2 K - 273 K
Authors:R. Gerald  L.G. Hyman  R. Ladbury  R. Rezmer  E. Fernandez
Affiliation:The authors are at the Argonne National Laboratory, Argonne, IL 60439, USA
Abstract:We describe how a simple two-point calibration of a platinum resistance thermometer can be used to deduce the resistance-temperature relationship in the range 4.2 K ? T ? 273 K. This relationship is established by comparing the unknown thermometer to two carefully calibrated platinum resistance thermometers. This technique produces an accuracy of (0.1 – 0.3) K. This accuracy should be adequate for many engineering applications.
Keywords:cryogenic  thermometer  calibration
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏