Effects of Crimping on Mechanical Performance of Nitinol Stent Designed for Femoral Artery: Finite Element Analysis |
| |
Authors: | F Nematzadeh S K Sadrnezhaad |
| |
Affiliation: | 1. Department of Materials Engineering, Faculty of Engineering, Arak University, 38156-88349, Arak, Iran 2. Materials and Energy Research Center (MERC), 14155-4777, Tehran, Iran 3. Department of Materials Science and Engineering, Sharif University of Technology, 11155-9466, Tehran, Iran
|
| |
Abstract: | Nitinol stents are used to minimize improper dynamic behavior, low twistability, and inadequate radial mechanical strength of femoral artery stents. In this study, finite element method is used to investigate the effect of crimping and Austenite finish temperature (A f) of Nitinol on mechanical performance of Z-shaped open-cell femoral stent under crimping conditions. Results show that low A f Nitinol has better mechanical and clinical performance due to small chronic outward force, large radial resistive force, and appropriate superelastic behavior. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|