首页 | 本学科首页   官方微博 | 高级检索  
     


Electronically swept millimeter-wave interferometer for spatially resolved measurement of plasma electron density
Authors:Howard John  Oliver David
Affiliation:Plasma Research Laboratory, Research School of Physical Sciences and Engineering, Australian National University, Canberra, ACT, Australia. john.howard@anu.edu.au
Abstract:We report the development and initial implementation of what we believe to be a new rapid- spatial-scan millimeter-wave interferometer for plasma density measurements. The fast scan is effected by electronic frequency sweeping of a wideband (180-280 GHz) backward-wave oscillator whose output is focused onto a fixed blazed diffraction grating. The system, which augments the rotating-grating scanned multiview H-1 heliac interferometer, can sweep the plasma cross section in a period of less than 1 ms with a beam diameter in the plasma of 20 mm and phase noise of the order of 0.01 rad.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号