首页 | 本学科首页   官方微博 | 高级检索  
     

基于阈值增量的标准动态眩光源设计方法
引用本文:乔卫东,张恒,王嘉明,李奕.基于阈值增量的标准动态眩光源设计方法[J].计量学报,2022,43(3):338-345.
作者姓名:乔卫东  张恒  王嘉明  李奕
作者单位:1.西安理工大学 机械与精密仪器工程学院,陕西 西安 710048
2.陕西省计量科学研究院 光学计量测试中心,陕西 西安 710065
基金项目:国家重点研发计划(2018YFF0212105);
摘    要:针对目前眩光测量装置面临的测量标准不统一的问题,提出了一种基于阈值增量(TI)的标准动态眩光源设计方法,建立了眩光测量仪校准系统.依据阈值增量的测量要求确定标准动态眩光源结构;眩光测量过程中,在保证实验室环境和现实路灯照明环境中的位置指数与立体角分别相等的前提下,规划眩光源的布局以及确定眩光源的发光面积;以最高为24 ...

关 键 词:计量学  动态眩光源  阈值增量  位置指数  不确定度
收稿时间:2020-12-10

Standard Dynamic Glare Light Source Design Method Based on Threshold Increment
QIAO Wei-dong,ZHANG Heng,WANG Jia-ming,LI Yi.Standard Dynamic Glare Light Source Design Method Based on Threshold Increment[J].Acta Metrologica Sinica,2022,43(3):338-345.
Authors:QIAO Wei-dong  ZHANG Heng  WANG Jia-ming  LI Yi
Affiliation:1. School of Mechanical & Precision Instrument Engineering, Xi‘an University of Technology, Xi’an, Shaanxi 710048, China
2. Optical Metrology and Test Center, Shaanxi Institute of  Metrology Science, Xian, Shaanxi 710065, China
Abstract:Aiming at the problem of inconsistent measurement standards faced by current glare measuring devices, a standard dynamic glare source design method based on threshold increments (TI) is proposed, and a glare measuring instrument calibration system is established. The structure of the standard dynamic glare light source is determined according to the measurement requirements of the threshold increment. In the process of glare measurement, the layout of the glare source is planned and the luminescence of the glare source area is determined under the premise that the position index and solid angle in the laboratory environment and the real street lighting environment are equal respectively. Simulation is performed for the same glare value in different scenes and different glare values in different scenes with a frequency of up to 24Hz to match the glare situation in various environments. The experimental results show that the standard dynamic glare source covers 8.0%~26.7%, and its flicker frequency can be adjusted up to 24 Hz, and the extended uncertainty of simulated glare is 1.3%, which can meet the calibration requirements of various glare measuring instruments.
Keywords:metrology  dynamic glare light source  threshold increment  position index  uncertainty  
本文献已被 万方数据 等数据库收录!
点击此处可从《计量学报》浏览原始摘要信息
点击此处可从《计量学报》下载全文