Improved lateral force calibration based on the angle conversion factor in atomic force microscopy |
| |
Authors: | DUKHYUN CHOI,WOONBONG HWANG,& EUISUNG YOON&dagger |
| |
Affiliation: | Department of Mechanical Engineering, Pohang University of Science and Technology, San 31, Hyoja Dong, Nam-gu, Pohang, Kyungbuk 790-784, Korea;Tribology Laboratory of Korea Institute of Science and Technology, Seoul, Korea |
| |
Abstract: | ![]() A novel calibration method is proposed for determining lateral forces in atomic force microscopy (AFM), by introducing an angle conversion factor, which is defined as the ratio of the twist angle of a cantilever to the corresponding lateral signal. This factor greatly simplifies the calibration procedures. Once the angle conversion factor is determined in AFM, the lateral force calibration factors of any rectangular cantilever can be obtained by simple computation without further experiments. To determine the angle conversion factor, this study focuses on the determination of the twist angle of a cantilever during lateral force calibration in AFM. Since the twist angle of a cantilever cannot be directly measured in AFM, the angles are obtained by means of the moment balance equations between a rectangular AFM cantilever and a simple commercially available step grating. To eliminate the effect of the adhesive force, the gradients of the lateral signals and the twist angles as a function of normal force are used in calculating the angle conversion factor. To verify reliability and reproducibility of the method, two step gratings with different heights and two different rectangular cantilevers were used in lateral force calibration in AFM. The results showed good agreement, to within 10%. This method was validated by comparing the coefficient of friction of mica so determined with values in the literature. |
| |
Keywords: | Angle conversion factor atomic force microscopy coefficient of friction lateral force calibration lateral force calibration factor mica twist angle |
|
|