Upscaling DC resistance using a Josephson junction array |
| |
Authors: | Kohlmann J. Gutmann P. Niemeyer J. |
| |
Affiliation: | Phys.-Tech. Bundesanstalt, Braunschweig; |
| |
Abstract: | ![]() A ratio standard for DC resistance measurements in the range above about 10 kΩ using a three-terminal Josephson junction series array is described. Resistance ratios of two-terminal resistors from 1:1 to at least 1:10 are examined with a single device utilizing the voltage steps of the Josephson array as voltage sources of a potentiometer bridge configuration. Some features of the circuit are discussed. The system promises systematic ratio errors below 10-8 and uncertainties which depend only on the quality of the resistors. Preliminary measurements with a prototype instrument show a relative reproducibility of 1×10-7, limited by the noise of a commercial digital multimeter used as a detector |
| |