首页 | 本学科首页   官方微博 | 高级检索  
     


Automated detection of ocular alignment with binocular retinal birefringence scanning
Authors:Hunter David G  Shah Ankoor S  Sau Soma  Nassif Deborah  Guyton David L
Affiliation:Zanvyl Krieger Children's Eye Center at the Wilmer Institute, The Johns Hopkins University School of Medicine, Baltimore, Maryland 21287, USA. david.hunter@tch.harvard.edu
Abstract:We previously developed a retinal birefingence scanning (RBS) device to detect eye fixation. The purpose of this study was to determine whether a new binocular RBS (BRBS) instrument can detect simultaneous fixation of both eyes. Control (nonmyopic and myopic) and strabismic subjects were studied by use of BRBS at a fixation distance of 45 cm. Binocularity (the percentage of measurements with bilateral fixation) was determined from the BRBS output. All nonstrabismic subjects with good quality signals had binocularity >75%. Binocularity averaged 5% in four subjects with strabismus (range of 0-20%). BRBS may potentially be used to screen individuals for abnormal eye alignment.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号