Automated detection of ocular alignment with binocular retinal birefringence scanning |
| |
Authors: | Hunter David G Shah Ankoor S Sau Soma Nassif Deborah Guyton David L |
| |
Affiliation: | Zanvyl Krieger Children's Eye Center at the Wilmer Institute, The Johns Hopkins University School of Medicine, Baltimore, Maryland 21287, USA. david.hunter@tch.harvard.edu |
| |
Abstract: | We previously developed a retinal birefingence scanning (RBS) device to detect eye fixation. The purpose of this study was to determine whether a new binocular RBS (BRBS) instrument can detect simultaneous fixation of both eyes. Control (nonmyopic and myopic) and strabismic subjects were studied by use of BRBS at a fixation distance of 45 cm. Binocularity (the percentage of measurements with bilateral fixation) was determined from the BRBS output. All nonstrabismic subjects with good quality signals had binocularity >75%. Binocularity averaged 5% in four subjects with strabismus (range of 0-20%). BRBS may potentially be used to screen individuals for abnormal eye alignment. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |