A study of the accuracy of the Precision Q.I.D. glucometer
Authors:HA Usmani  II Khan  FH Mughal
Affiliation:Faculty of Health Sciences, Aga Khan University, Karachi.
Abstract:
The Precision Quantum-in-dium (Q.I.D.) Glucometer was used to determine the glucose concentrations of 38 human blood samples. The same samples were also run on the OPERA Chemistry Autoanalyzer 2010 system. More than 44% of the glucometer reports had a difference of greater than 15% from the respective autoanalyzer reports. The calibration of the glucometer could be the source of the error and an improvement is recommended.
Keywords: