首页 | 本学科首页   官方微博 | 高级检索  
     


High-performance liquid chromatography-atmospheric pressure photoionization/tandem mass spectrometric analysis for small molecules in plasma
Authors:Hsieh Yunsheng  Merkle Kara  Wang Ganfeng  Brisson Jean-Marc  Korfmacher Walter A
Affiliation:Drug Metabolism and Pharmacokinetics Department, Schering-Plough Research Institute, Kenilworth, New Jersey 07033, USA. yunsheng.hsieh@spcorp.com
Abstract:A generic high-performance liquid chromatography (HPLC) system interfaced with an atmospheric pressure photoionization (APPI) source and a tandem mass spectrometer was developed for the quantitative determination of small molecules in plasma in support of exploratory in vivo pharmacokinetics. This report summarizes the effects of variations in reversed-phase mode HPLC conditions such as mobile-phase flow rate, solvent composition, organic modifier content, and nebulizer temperature on the photoionization efficiency of both clozapine and lonafarnib. The matrix ionization suppression effect on this method was investigated using the postcolumn infusion technique. The procedure was used to quantitate plasma levels following oral administration of 42 drug discovery compounds to rats. The pharmacokinetic results of 42 drug discovery compounds in rats evaluated by both APPI and atmospheric pressure chemical ionization interfaces were found to be well correlated.
Keywords:
本文献已被 PubMed 等数据库收录!