首页 | 本学科首页   官方微博 | 高级检索  
     


Characteristics of a prototype CdZnTe detector
Authors:XU Huichao   CHENG Cheng   ZHAO Cuilan   ZHANG Jinzhou  PU Shijie
Affiliation:[1]Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai 201800, China [2]No.3 Research Institute of Ministry of Public Security, Shanghai 200031,China
Abstract:Cadmium zinc telluride(CZT)is a preferred material for X-ray and gamma-ray detector.Thanks to the relatively high atomic number,high density and wide band-gap.CZT detector possesses sharp energy resolution and high detection efficiency without cryogenic cooling.We have developed a CdZnTe detector with an energy resolution of 3.45%(FWHM)at 59.54 keV at room temperature.and it is used for X-ray fluorescence analysis.In this paper,leakage current,energy resolution and long term stability of the CZT detector are discussed.
Keywords:CDZnTe(CZT)detector  Fabrication  Characteristics  Stability  detector  CdZnTe  prototype  energy resolution  term  stability  leakage  current  paper  used  fluorescence analysis  room temperature  FWHM  detection efficiency  cooling  density  wide  high  atomic number  material
本文献已被 CNKI 维普 万方数据 ScienceDirect 等数据库收录!