首页 | 本学科首页   官方微博 | 高级检索  
     


Focused Array Hyperthermia Applicator: Theory and Experiment
Authors:Gee  Walter Lee  Shung-Wu Bong  Nelson K Cain  Charles A Mittra  Raj Magin  Richard L
Affiliation:University of Illinois, Urbana, IL 61801 and Lockheed Missiles and Space Co., Inc.;
Abstract:The design and analysis of a focused linear array at 2450 MHz for microwave hyperthermia research is described. The array, which was submerged in deionized water to reduce its size and to provide a better impedance match to a high dielectric medium representative of human tissues, consisted of four titanium dioxide loaded horn antennas with apertures of 2.0 x 1.4 cm and a feed network with weighted phase shifts, Power radiation pattern measurements were made in planes ranging from 7.6 to 10.2 cm from the array to determine the focusing characteristics and beam spot size. Due to high attenuation in the medium, planes beyond 10.2 cm were not considered. The half-power beamwidth (HPBW) measured at the focal point was approximately 1.3 cm. The measured patterns were found to be in close agreement with theoretical predictions.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号