Using LabVIEW to facilitate calibration and verification for respiratory impedance plethysmography. |
| |
Authors: | W S Ellis R T Jones |
| |
Affiliation: | Langley Porter Psychiatric Institute, University of California, San Francisco 94143. |
| |
Abstract: | A system for calibrating the Respitrace impedance plethysmograph was developed with the capacity to quantitatively verify the accuracy of calibration. LabVIEW software was used on a Macintosh II computer to create a user-friendly environment, with the added benefit of reducing development time. The system developed enabled a research assistant to calibrate the Respitrace within 15 min while achieving an accuracy within the normally accepted 10% deviation when the Respitrace output is compared to a water spirometer standard. The system and methods described were successfully used in a study of 10 subjects smoking cigarettes containing marijuana or cocaine under four conditions, calibrating all subjects to 10% accuracy within 15 min. |
| |