首页 | 本学科首页   官方微博 | 高级检索  
     


Simulation and optimization of a piezoelectric micropump for medical applications
Authors:Qifeng Cui  Chengliang Liu  Xuan F Zha
Affiliation:(1) School of Mechanical Engineering, Shanghai JiaoTong University, Dongchuan Road 800, Shanghai, 200240, China;(2) National Institute of Standards and Technology (NIST) and the University of Maryland, Gaithersburg, MD 20899, USA
Abstract:We designed a valveless micropump excited by a piezoelectric actuator for medical applications. The complete electric–fluid–solid coupling model is built upon using ANSYS software (Canonsburg, PA) to investigate the behaviors of the micropump. The effects of the geometrical dimensions on the micropump characteristics and its efficiency are analyzed. The simulation results show that there is an optimal thickness of the piezoelectric layer to obtain a large pump flow, and that this optimal thickness is affected by the material and the thickness of the pump membrane. To enhance the performance of the micropump, some important diffuser parameters, such as the diffuser length, the diffuser angle, and the neck width, should be optimized. However, the variations of the diffuser’s geometrical dimensions do not affect the optimal thickness of the piezoelectric layer.
Keywords:MEMS  Valveless micropump  Medical application  Piezoelectric actuation  Simulation  Optimization
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏