首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of Average Ion Temperature of the TVD Tokamak by a Simple Charge Exchange Analyzer Detector
Authors:M. Kazemi  M. Habibi  R. Amrollahi
Affiliation:1.Energy Engineering and Physics Department,Amirkabir University of Technology,Tehran,Iran
Abstract:In this paper the average ion temperature of the TVD tokamak plasma is determined by a low cost charge exchange analyzer. When the neutral beam energy was equal to 3 keV, the stripping efficiency was 0.065. Estimated average ion temperatures for cold and hot regions of the plasma were 100 and 650 eV, respectively. Unlike the cold region of the plasma in which ion temperature was not related to the time, ion temperature in the hot region was maximum when the plasma current reached its maximum value.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏