Abstract: | On-line processing and analysis of high resolution electron microscope (HREM) images has been implemented using a software controlled image processor based on a commercial digital video framestore. It directly digitizes the output from the low light TV camera on the HREM and applies any processing necessary. The conditions under which images from amorphous specimens are being obtained can be established on-line. The fast fourier transforms (FFTs) of 256 times 256 pixels produced in 15 s with the system off-line are comparable to those from diffraction analysis of the same data on a laser optical bench. The precision of analysis of defocus, aberrations and astigmatism on-line is discussed, together with possible approaches to routine analysis of crystalline specimens and for interactive control. |