首页 | 本学科首页   官方微博 | 高级检索  
     


A Time of Flight-Energy spectrometer for stopping power measurements in Heavy Ion-ERD analysis at iThemba LABS
Authors:M. Msimanga  C.M. Comrie  S. Murray  G. Dollinger
Affiliation:a iThemba LABS, P.O. Box 722, Somerset West 7129, Cape Town, South Africa
b University of Cape Town, Rondebosch 7701, South Africa
c Faculty of Health and Wellness Sciences, CPUT, Cape Town, South Africa
d TU München, Physik Department E 12, D-85748 Garching, Germany
Abstract:The quantitative analysis of thin layers using Heavy Ion-Elastic Recoil Detection (HI-ERD) can be reliably performed if the stopping powers of the probing ions and recoils in a given target matrix are known accurately. Unfortunately for many projectile/target combinations experimental data is limited and where available, deviations of up to 50% between experiment and theory have been reported. This presentation describes the assembly of a Time of Flight-Energy (ToF-E) detector system developed for HI-ERD analysis and adapted for stopping power measurements at iThemba LABS. First results from energy loss measurements of 0.1-0.5 MeV/nucleon 28Si and 84Kr ions in ZrO2 are presented and compared with predictions of the widely used SRIM2003 (Stopping Range of Ions in Matter).
Keywords:34.50.Bw   82.80.Yc
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号