用150 W氘灯标定200~300 nm光谱辐照度
引用本文:黄煜,王淑荣,张振铎,林冠宇,李福田.用150 W氘灯标定200~300 nm光谱辐照度[J].光学精密工程,2007,15(8):1215-1219.
作者姓名:黄煜  王淑荣  张振铎  林冠宇  李福田
作者单位:1. 中国科学院,长春光学精密机械与物理研究所,应用光学国家重点实验室,吉林,长春,130033;中国科学院,研究生院,北京,100039
2. 中国科学院,长春光学精密机械与物理研究所,应用光学国家重点实验室,吉林,长春,130033
摘    要:采用150 W大功率氘灯标定光谱仪器的紫外波段光谱辐照度响应度,解决了大多数光谱辐射计由于在紫外波段响应度低、信噪比小导致的定标困难。在此基础上以氘灯加球面反射镜构建了平行光光谱辐照度定标单元,满足了某些遥感类光谱辐射计辐照度响应度定标要求,该单元通过照度标准传递的方式获得标准光谱辐照度值,避开了球面反射镜光谱反射率的测量难点。与美国NIST 的FEL标准灯标定同一台光谱辐射计,其结果最大偏差为3%,200~300 nm定标误差为4.7%。

关 键 词:氘灯  光谱辐射计  照度传递  照度定标
文章编号:1004-924X(2007)08-1215-05
收稿时间:2007-05-21
修稿时间:2007-05-212007-06-13

Calibration of 200~300 nm spectral irradiance using 150 W deuterium lamp
HUANG Yu,WANG Shu-rong,ZHANG Zhen-duo,LIN Guan-yu,LI Fu-tian.Calibration of 200~300 nm spectral irradiance using 150 W deuterium lamp[J].Optics and Precision Engineering,2007,15(8):1215-1219.
Authors:HUANG Yu  WANG Shu-rong  ZHANG Zhen-duo  LIN Guan-yu  LI Fu-tian
Affiliation:1. State Key Laboratory of Applied Optics , Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China ; 2. Graduate School of the Chinese Academy of Sciences,Beijing 100039, China
Abstract:A calibrating method for the irradiance responsivity of 200~300 nm spectral instruments using 150 W deuterium was introduced in detail. This method overcame the troubles caused by the weak ultraviolet response of most spectral radiometer and poor signal-to-noise ratio. A spectral irradiance calibration unit based on parallel light was established to meet the calibration conditions of some remote sensing spectral radiometers. The calibration unit got the standard value of spectral irradiance by irradiance transfer, which avoided the difficulty of spectral reflectance measurement of the concave mirror. Using the unit and a FEL lamp from NIST to calibrate the same ultraviolet radiometer, the results show that the bias between them is 3%, and the error of spectral irradiance calibration is 4.7% in range 200~300 nm.
Keywords:deuterium lamp  spectral radiometer  irradiance transfer  irradiance calibration