On the Nature and Elimination of Stimulus Artifact in Nerve Signals Evoked and Recorded Using Surface Electrodes |
| |
Authors: | McGill Kevin C Cummins Kenneth L Dorfman Leslie J Berlizot Bruno B Luetkemeyer Kelly Nishimura Dwight G Widrow Bernard |
| |
Affiliation: | Department of Electrical Engineering, Stanford University; |
| |
Abstract: | The electrical stimulus pulse and the surface electrodes commonly used to study compound action potentials of peripheral nerves give rise to an artifact consisting of an initial spike and a longer lasting tail which often interferes with the recorded signal. The artifact has four sources: 1) the voltage gradient between the recording electordes caused by stimulus current flowing through the limb, 2) the common-mode voltage of the limb caused by current escaping through the ground electrode, 3) the capacitive coupling between the stimulating and recording leads, and 4) the high-pass filtering characteristics of the recording amplifier. This paper models these sources and presents several methodological rules for minimizing their effects. Also presented are three computer-based methods for subtracting the residual artifact from contaminated records using estimates of the artifact obtained from: 1) subthreshold stimulation, 2) a second recording site remote from the nerve, or 3) stimulation during the refractory period of the nerve. |
| |
Keywords: | |
|
|