UV-LIGA interferometer biosensor based on the SU-8 optical waveguide
Authors:BY  CH  YC  YH
Affiliation:

aDevice Technology Group, National Synchrotron Radiation Research Center, Hsinchu 30077, Taiwan

bDepartment of Mechanical Engineering, National Chiao-Tung University, Hsinchu 30077, Taiwan

Abstract:SU-8 resist was used as a core/cladding waveguide material to fabricate a Mach–Zehnder interferometer (MZI) for biochemical sensing. The refractive index of the SU-8 resist was fine-tuned with a Δn of 0.004 for single-mode transmission. The UV lithography processes of the SU-8 resist were also optimized to pattern the high resolution (1 μm) and high aspect ratio (AR = 6) Y-branch structure of the optical interferometer. Optical measurements reveal that the SU-8 MZI chip can efficiently transmit the NIR laser (λ = 1310 nm) with a total loss less than 6 dB. When one branch of the MZI is in contact with the analyte, the interfered intensity stabilizes after the soaking time exceeds 5 min. NaCl solution with a concentration of 10−9 g/l can be detected using the SU-8 MZI chip. In the future, the polymer MZI chip can be mass-produced by molding process (or the LIGA process). The low-cost, label-free, real-time and high-sensitivity MZI chip will benefit many applications related to biological, environmental and industrial detection.
Keywords:Optical waveguide  Interferometer  Biochemical sensor  SU-8 resist  LIGA process