首页 | 本学科首页   官方微博 | 高级检索  
     


Axial strain calculation using a low-pass digital differentiator in ultrasound elastography
Authors:Luo Jianwen  Bai Jing  He Ping  Ying Kui
Affiliation:Department of Biomedical Engineering, Tsinghua University, Beijing 100084, PR China. deabj@tsinghua.edu.cn
Abstract:In ultrasound elastography, tissue axial strains are calculated from the gradient of the estimated axial displacements. However, the common differentiation operation amplifies the noises in the displacement estimation, especially at high frequencies. In this paper, a low-pass digital differentiator (LPDD) is proposed to calculate the axial strain from the estimated tissue displacement. Several LPDDs that have been well developed in the field of digital signal processing are presented. The corresponding performances are compared qualitatively and quantitatively in computer simulations and in preliminary phantom and in vitro experiments. The results are consistent with the theoretical analysis of the LPDDs.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏