SoI--gel modification of pH electrode glass membranes for sensing anions and metal ions |
| |
Authors: | Kimura K Yajima S Takase H Yokoyama M Sakurai Y |
| |
Affiliation: | Department of Applied Chemistry, Faculty of Systems Engineering, Wakayama University, Japan. kimura@sys.wakayama-u.ac.jp |
| |
Abstract: | To obtain glass membrane electrodes selective for anions and metal ions, pH electrode glass membranes were modified by a sol-gel method using a quaternary ammonium salt and a bis(crown ether). A chloride ion-sensing glass membrane was designed, in which a pH electrode glass membrane was modified chemically by an alkoxysilyl quaternary ammonium chloride. X-ray photoelectron spectroscopy confirmed the chemical bonding of the quaternary ammonium moiety to the starting glass surface, which afforded the first example of glass-based "anion"-sensing membranes. A neutral carrier-type sodium ion-selective glass membrane was also fabricated which encapsulates a bis(12-crown-4) derivative in its sol-gel-derived surface. Both sol-gel-modified anion and metal ion-selective glass electrodes exhibited high sensitivity to their ion activity changes. The present sol-gel modification paves the way for designing glass-based ion sensors with tailor-made ion selectivities toward anions as well as cations. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|