首页 | 本学科首页   官方微博 | 高级检索  
     


Atomic force microscope cantilever calibration using a focused ion beam
Authors:Slattery Ashley D  Quinton Jamie S  Gibson Christopher T
Affiliation:Smart Surface Structures Group, Flinders Centre for NanoScale Science and Technology, School of Chemical and Physical Sciences, Flinders University, Bedford Park, SA 5042, Australia.
Abstract:A calibration method is presented for determining the spring constant of atomic force microscope (AFM) cantilevers, which is a modification of the established Cleveland added mass technique. A focused ion beam (FIB) is used to remove a well-defined volume from a cantilever with known density, substantially reducing the uncertainty usually present in the added mass method. The technique can be applied to any type of AFM cantilever; but for the lowest uncertainty it is best applied to silicon cantilevers with spring constants above 0.7?N?m(-1), where uncertainty is demonstrated to be typically between 7 and 10%. Despite the removal of mass from the cantilever, the calibration method presented does not impair the probes' ability to acquire data. The technique has been extensively tested in order to verify the underlying assumptions in the method. This method was compared to a number of other calibration methods and practical improvements to some of these techniques were developed, as well as important insights into the behavior of FIB modified cantilevers. These results will prove useful to research groups concerned with the application of microcantilevers to nanoscience, in particular for cases where maintaining pristine AFM tip condition is critical.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号