首页 | 本学科首页   官方微博 | 高级检索  
     


The n-PI-method for helical cone-beam CT
Authors:Proksa R  Köhler T  Grass M  Timmer J
Affiliation:Philips Research Laboratory, Division Technical Systems Hamburg, Germany.
Abstract:A new class of acquisition schemes for helical cone-beam computed tomography (CB-CT) scanning is introduced, and their effect on the reconstruction methods is analyzed. These acquisition schemes are based on a new detector shape that is bounded by the helix. It will be shown that the data acquired with these schemes are compatible with exact reconstruction methods, and the adaptation of exact reconstruction algorithms to the new acquisition geometry is described. At the same time, the so-called PI-sufficiency condition is fulfilled. Moreover, a good fit to the acquisition requirements of the various medical applications of cone-beam CT is achieved. In contrast to other helical cone-beam acquisition and reconstruction methods, the n-PI-method introduced in this publication allows for variable pitches of the acquisition helix. This additional feature will introduce a higher flexibility into the acquisition protocols of future medical cone-beam scanners. An approximative n-PI-filtered backprojection (n-PI-FBP) reconstruction method is presented and verified. It yields convincing image quality.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号