首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of three commercially available menual "kit" methods for triglycerides in serum
Authors:SL Kykstra  JD Lawrence
Abstract:
We compared three commercially available kits for use in determining triglycerides (triacylglycerols) in serum, by different methods, for accuracy and precision, with use of pooled sera of low, normal, and high triglyceride concentrations. Accuracy varied from -17% to +16% of the independently determined value, depending on the method and triglyceride concentration. In general, the Calbiochem method was the most nearly accurate, and only it met Center for Disease Control criteria for precision at all three triglyceride concentrations. The relationships of mean values to the suggested normal range for each method varied, some significant differences being found. Our data demonstrate the importance of a correlation between the independently determined value and the normal range for the method used and stresses the need for normal-range values to accompany all patients' reports.
Keywords:
本文献已被 PubMed 等数据库收录!