Image quality and dose in spiral computed tomography |
| |
Authors: | FR Verdun RA Meuli FO Bochud C Imsand S Raimondi P Schnyder JF Valley |
| |
Affiliation: | Institut de Radiophysique Appliquée, Centre Universitaire, Lausanne, Switzerland. |
| |
Abstract: | Image quality and dose produced by a spiral CT with various pitch values were investigated on the basis of test-object images and measurements of CT dose indexes. No major difference in image quality or dose was observed when comparing sequential and spiral mode acquisition with a pitch of one. Increase in pitch value produces a loss of contrast which leads to a loss of detectability. Nevertheless, in realistic protocols the image quality in the transverse plane remains acceptable up to pitch values of 1.3-1.6. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|