首页 | 本学科首页   官方微博 | 高级检索  
     


Artefacts in magnetic resonance imaging caused by dental material
Authors:Georg Eggers  Marcus Rieker  Bodo Kress  Jochen Fiebach  Hartmut Dickhaus  Stefan Hassfeld
Affiliation:(1) Department of Oral and Cranio-Maxillofacial Surgery, Heidelberg University Hospital, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany;(2) Department of Neuroradiology, Heidelberg University Hospital,;(3) Institute for Medical Biometry and Informatics, Heidelberg University,
Abstract:
A common problem in computer tomography (CT) based imaging of the oral cavity is artefacts caused by dental restorations. The aim of this study was to investigate whether magnetic resonance imaging (MRI) of the oral cavity would be less affected than CT by artefacts caused by typical dental restorative alloys. In order to assess the extent of artefact generation, corresponding MRI scans of the same anatomic region with and without dental metal restorations were matched using a stereotactic frame. MRI imaging of the oral and maxillofacial region could be performed without reduction of the image quality by metallic dental restorations made from titanium, gold or amalgam. Dental restorations made from titanium, gold or amalgam did not reduce the image quality of the MRI sequence used in imaging of the oral and maxillofacial region for dental implant planning. In this respect MRI is superior to CT in implant planning.
Keywords:MR tomography  Artefacts  Dental material
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号