首页 | 本学科首页   官方微博 | 高级检索  
     


Reproducibility of color Doppler imaging in orbital vessels
Authors:BC Senn  HJ Kaiser  A Sch?tzau  J Flammer
Affiliation:Medizinische Hochschule Hannover, Abteilung Gastroenterologie und Hepatologie.
Abstract:The purpose of this study was to evaluate the short-term reproducibility, the long-term reproducibility, and the reproducibility of the measurements made between two different examiners of blood-flow velocity in extraocular vessels using color Doppler imaging (CDI). In a group of 10 healthy volunteers, measurements of the peak-systolic velocity (PSV), end-diastolic velocity (EDV), and resistivity index (RI) in the ophthalmic artery and the central retinal artery as well as one lateral and one medial short posterior ciliary artery were performed. The best intraindividual reproducibility, expressed as the relative error, was found for the PSV and RI measured in the ophthalmic artery (OA; PSV 5.9%, RI 3.1%) and the central retinal artery (CRA; PSV 7.7%, RI 4.7%). The PSV and RI determined in the posterior ciliary arteries were less reproducible but as much so as the EDV measured in the OA (11.8%) and CRA (19.9%). No systematic trend could be found between the first and second measurements. However, the long-term fluctuation was considerable high. Measurements of PSVs showed good concurrence between the two observers in all four vessels measured. The EDVs and RIs differed statistically significantly between the two observers. This difference was most pronounced in the posterior ciliary arteries.
Keywords:
本文献已被 PubMed 等数据库收录!