首页 | 本学科首页   官方微博 | 高级检索  
     

基于四极杆-线形离子阱串联质谱技术测量血清25OHD
引用本文:谢洁,易可可,高文静,刘浩,李家练,黄泽建,刘梅英,余青霓,王涵文,江游,戴新华,方向.基于四极杆-线形离子阱串联质谱技术测量血清25OHD[J].质谱学报,2023,44(1):13-24.
作者姓名:谢洁  易可可  高文静  刘浩  李家练  黄泽建  刘梅英  余青霓  王涵文  江游  戴新华  方向
作者单位:中国计量科学研究院前沿计量科学中心,国家市场监管技术创新中心(质谱),北京100029;常州磐诺仪器有限公司,江苏 常州213163;安徽砺沐医疗器械有限公司,安徽 六安237001
基金项目:国家重点研发计划——重大科学仪器设备开发专项项目(2020YFF01014603);;国家自然科学基金——国家重大科研仪器研制项目(21927812);
摘    要:本研究基于自主研制的QLIT-6610MD四极杆-线形离子阱液相色谱-质谱联用仪,建立了准确测量血清25OHD的同位素稀释质谱法。血清样品用甲醇沉淀蛋白,正己烷液-液萃取,上清液用氮气吹干,初始流动相复溶,QLIT-6610MD测定。采用基质匹配标准曲线定量,25OHD2和25OHD3分别在0.78~50μg/L和1.56~100μg/L范围内的线性决定系数R2>0.999 9,检出限和定量限为0.06、0.12μg/L和0.02、0.06μg/L,质控品精密度优于4.06%,血清样本添加回收率为99.25%~108.38%,变异系数≤5.10%。用QLIT-6610MD、AB 4500MD和YS EXACT 9900MD 3台液相色谱-质谱联用仪测定50例临床血清,采用Pearson系数和Bland-Altman法分析,3组数据的相关系数R≥0.998 0、P<0.01。这表明3台仪器所测结果一致,QLIT-6610MD可以达到三重四极杆质谱仪定量测定血清25OHD的水平,能够满足临床诊断需求,将为临床质谱...

关 键 词:液相色谱-串联质谱(LC-MS/MS)  同位素稀释质谱  四极杆  线形离子阱  25羟基维生素D(25OHD)

Determination of 25OHD in Serum Based on Quadrupole-Linear Ion Trap Tandem Mass Spectrometric Technique
XIE Jie,YI Ke-ke,GAO Wen-jing,LIU Hao,LI Jia-lian,HUANG Ze-jian,LIU Mei-ying,YU Qing-ni,WANG Han-wen,JIANG You,DAI Xin-hua,FANG Xiang.Determination of 25OHD in Serum Based on Quadrupole-Linear Ion Trap Tandem Mass Spectrometric Technique[J].Journal of Chinese Mass Spectrometry Society,2023,44(1):13-24.
Authors:XIE Jie  YI Ke-ke  GAO Wen-jing  LIU Hao  LI Jia-lian  HUANG Ze-jian  LIU Mei-ying  YU Qing-ni  WANG Han-wen  JIANG You  DAI Xin-hua  FANG Xiang
Affiliation:Technology Innovation Center of Mass Spectrometry for State Market Regulation, Center for Advanced Measurement Science, National Institute of Metrology, Beijing 100029, China;Changzhou Panna Instrument Co., Ltd., Changzhou 213163, China;Anhui Limu Medical Instrument Co., Ltd., Liuan 237001, China
Abstract:Vitamin D deficiency and insufficiency is a global health issue that afflicts people of different age groups all over the world. As a reliable indicator for the evaluation of vitamin D, 25OHD is usually determined by LC-MS/MS. The tandem mass spectrometry based on ion trap has the characteristics of simple structure and high sensitivity. However, there are some problems such as narrow linear range and space charge effect, which limit its analysis of trace substances in complex matrix samples. In this study, the device and technology of a new quadrupole linear ion trap mass spectrometry was used to reduce the space charge interference of trace or ultra trace target ions by high abundance impurity ions in complex matrices. For the first time, isotope dilution mass spectrometry for accurate measurement of serum 25OHD was established by QLIT-6610MD, which is quadrupole-linear ion trap tandem mass spectrometer, and was independently developed by Technology Innovation Center of Mass Spectrometry for State Market Regulation, Center for Advanced Measurement Science, National Institute of Metrology. Serum samples were precipitated with methanol, extracted with n-hexane, dried with liquid nitrogen, redissolved in the initial mobile phase, and determined by QLIT-6610MD. The matrix-matched standard curve was used for quantification. The linear ranges of 25OHD2 and 25OHD3 were in the range of 0.78-50 μg/L and 1.56-100 μg/L with R2>0.999 9, respectively. The detection and quantification limits of 25OHD2 and 25OHD3 were 0.06, 0.12 μg/L and 0.02, 0.06 μg/L, respectively. The precision of quality control was better than 4.06%. The recoveries ranged from 99.25% to 108.38%, and the coefficients of variation were all less than 5.1%. After the method confirmation, the quantitative results of clinical samples were compared with triple quadrupole tandem mass spectrometry. 50 Clinical serum samples were measured and compared by QLIT-6610MD, AB 4500MD and YS EXACT 9900MD. The correlation and consistency of the data were analyzed by Pearson and Bland-Altman. The correlation coefficient of the three groups data was R≥0.998 0, P<0.01, and the results of Bland-Altman analysis showed good consistency, indicating that the results measured by the three instruments were consistent, which meant that QLIT-6610MD can meet the needs of clinical diagnosis. QLIT-6610MD achieved the same level of triple quadrupole mass spectrometry to determine the concentration of 25OHD in serum. It can not only become a new choice for clinical mass spectrometry, but also lay a foundation for the development of domestic mass spectrometer. Based on the direct axial coupling technology of quadrupole and linear ion trap, QLIT-6610MD can greatly reduce the interference of matrix ions and make the ion trap detect target ions more effectively. This technology has a strong potential to further improve the measurement accuracy of trace or ultra trace targets in complex matrices. In the future, it can be applied to mass spectrometry monitoring scenarios that needs to meet the requirements of high accuracy, high sensitivity, on-site speed and low cost at the same time.
Keywords:liquid chromatography tandem mass spectrometry (LC-MS/MS)  isotope dilution mass spectrometry  quadrupole  linear ion trap  25OHD  
点击此处可从《质谱学报》浏览原始摘要信息
点击此处可从《质谱学报》下载全文